標題: 人工膝關節所用超高分子量聚乙烯磨耗分析
The wear analysis of ultra high molecular weight polyethylene in artificial knee joint
作者: Chan-Hsien Chen
陳政賢
指導教授: 蘇侃
關鍵字: 人工膝關節,超高分子量聚乙烯,磨耗,
artificial knee joint,wear,polyethylene,UHMWPE,
出版年 : 2006
學位: 碩士
摘要: 本文探討人工膝關節的磨耗型態,以試驗機模擬人體一個步態週期中人工膝關節磨耗情形,包括其正向受力,前後移動及彎曲伸直的各項運動。在本文中採用兩種不同接觸壓應力大小作磨耗試驗,分別為16.5MPa和9.76MPa。
實驗結果顯示在較大應力作用下,在前半段區域由於受到高應力及滑動的影響,產生明顯的刮痕,其磨耗的深度也大於其他區域。而在縱剖面的影響也較為顯著,在應力較大的試片中,可以觀察到縱剖面的疲勞發生處會伴隨垂直表面的裂縫,在壓應力16.5MPa週期數100萬次以及200萬次之試片皆可觀察到有貫穿試片的垂直裂縫。在9.76MPa壓應力作用下的試片,表面磨耗型態皆較為平坦,前半段區域並無明顯刮痕,整體的磨耗深度也較小,在縱剖面的觀察下,垂直表面的裂縫也大大減少,說明由於壓應力的減少,可以改變磨屑型態,進而降低磨耗的速度。
臨床取下來的聚乙烯試片中,可以觀察到和人工膝關節磨耗試驗機所模擬之試片具有相似的磨耗型態,大體上和較高應力的試片結果較為類似,其因滑動產生的刮痕和片狀脫落的磨屑以及多層重疊的結果都和本文實驗結果相似。
The wear patterns of ultra high molecular weight polyethylene (UHMWPE) used in artificial knee joint were studied in this research. A knee joint simulator was used to simulate the vertical load, anterior-posterior translation, flexion-extension and internal-external rotation in a complete walking cycle. Two different contact compressive stresses of 16.5MPa and 9.76MPa were set in this experiment.
The experimental results show that obvious wear print would be observed in anterior area, it was caused by bearing high stress and relative sliding. The wear depth in this area was also deeper than that in other area. Subjected to the higher stress (i.e. 16.5MPa), fatigue cracks occurred in axial section perpendicular to the wear surface of the specimens subjected to 1x106 cycle and 2x106 cycle-tests. Subjected to the compressive stress of 9.76MPa, wear situation of the UHMWPE specimen was mild, and no obvious scrape in anterior area. The total wear depth was smaller, only some perpendicular cracks were observed.
The similar wear patterns were observed on both the clinically
used polyethylene and simulated one, especially on the specimens subjected to higher stress. The wear phenomena of delamination wear、stacks of wear debris and scratched marks on the simulated specimens were similar to those on the clinically used UHMWPE.
URI: http://tdr.lib.ntu.edu.tw/jspui/handle/123456789/34441
全文授權: 有償授權
顯示於系所單位:機械工程學系

文件中的檔案:
檔案 大小格式 
ntu-95-1.pdf
  目前未授權公開取用
13.22 MBAdobe PDF
顯示文件完整紀錄


系統中的文件,除了特別指名其著作權條款之外,均受到著作權保護,並且保留所有的權利。

社群連結
聯絡資訊
10617臺北市大安區羅斯福路四段1號
No.1 Sec.4, Roosevelt Rd., Taipei, Taiwan, R.O.C. 106
Tel: (02)33662353
Email: ntuetds@ntu.edu.tw
意見箱
相關連結
館藏目錄
國內圖書館整合查詢 MetaCat
臺大學術典藏 NTU Scholars
臺大圖書館數位典藏館
本站聲明
© NTU Library All Rights Reserved