請用此 Handle URI 來引用此文件:
http://tdr.lib.ntu.edu.tw/jspui/handle/123456789/27487
標題: | 以CMOS 0.18微米製程實現400 MHz植入式醫療通訊系統之數位校正超再生接收機 A 400-MHz Super-Regenerative Receiver with Digital Calibration for MICS Applications in 0.18-μm CMOS Process |
作者: | Ching-Jen Tung 董景仁 |
指導教授: | 林宗賢(Tsung-Hsien Lin) |
關鍵字: | 超再生接收機, super-regenerative receiver, |
出版年 : | 2007 |
學位: | 碩士 |
摘要: | In this work, we adopt the super-regenerative receiver to implement the receiver for the Medical Implant Communications Service (MICS) applications. The receiver is designed to demodulate the on-off keying signal. The architecture is simple and consumes low power. The proposed topology incorporates digital circuits to generate the quench signal on chip. In addition, an open-loop frequency calibration loop is proposed to adjust the digitally-controlled oscillator frequency to the MICS band. Although a phase-locked loop can also lock the oscillator frequency and achieve precise frequency calibration, its locking time is long. This proposed frequency calibration loop can adjust the oscillator frequency in 4 μs; thus, it saves energy consumption.
Fabricated in a 0.18-μm CMOS process, the measurement results show the proposed topology dissipates 3.5 mA from a 1.5-V supply. The total calibration time is less than 20 μs. The receiver has a sensitivity of -83 dBm and a maximum data rate of 1 Mbps. The chip area of the fully-integrated super-regenerative receiver is 1.47 mm X 1.46 mm including pads. The core area is 0.9 mm X 1 mm. |
URI: | http://tdr.lib.ntu.edu.tw/jspui/handle/123456789/27487 |
全文授權: | 有償授權 |
顯示於系所單位: | 電子工程學研究所 |
文件中的檔案:
檔案 | 大小 | 格式 | |
---|---|---|---|
ntu-96-1.pdf 目前未授權公開取用 | 2.56 MB | Adobe PDF |
系統中的文件,除了特別指名其著作權條款之外,均受到著作權保護,並且保留所有的權利。