標題: | 運用無線感測器網路於血液透析中心的病患生理值監測 Vital Signs Monitoring of Patients in a Hemodialysis Center Using Wireless Sensor Networks |
作者: | Yung-Cheng Wu 吳勇成 |
指導教授: | 江昭皚 |
關鍵字: | 透析中低血壓,血液透析,無線感測器網路,全球行動通訊系統,數據資料庫, Intradialytic hypotension,Hemodialysis,Wireless sensor networks,Global systems for mobile communications,MySQL database, |
出版年 : | 2009 |
學位: | 碩士 |
摘要: | 透析中低血壓(intradialytic hypotension, IDH)是透析治療中最常見的併發症,其發生的頻率大約有10-50%,老年人、糖尿病及心血管透析病患有更高的發生率,其會產生噁心、嘔吐及焦慮等症狀,嚴重甚至會出現休克和死亡的意外。加上血液透析中心護理人員因忙碌或交班,對病房病患生理監控易被臨床工作中斷,或因監測頻率不同、一人照顧多床病患,而影響醫療品質,因此,本研究整合無線感測器網路(Wireless Sensor Network, WSN)、全球行動通信系統(Global System for Mobile Communications, GSM)和數據資料庫,建立一個透析中心生理數據的自動化監測系統,將無線監測技術應用於病患透析中低血壓的產生,並且透過緊急通報系統,讓家屬或是照護人員即時得知透析中低血壓的發生,進而提供更即時的照護。
本系統適用範圍為醫院內血液透析中心的固定空間,系統架構主要分成兩大區塊,前端無線網路以及後端資訊處理主控平臺。在血液透析模擬實驗當中,我們模擬一固定空間裡的血液透析狀況,由六位受測者接受生理值監測,訂定時間區間為二小時連續測量,實驗結果顯示,生理值資料平均傳輸成功率為94.84%,生理值信度取樣分析結果顯示,收縮壓平均信度為1,舒張壓平均信度為0.89,而心搏率平均信度為0.94。實驗結果討論,生理值資料遺失率源於空間障礙物阻礙以及單向傳輸過程的資料碰撞。 Intradialytic hypotension (IDH) is a much common complication during a hemodialysis. The prevalence of IDH is about 10-50%. Elderly people, patients with diabetes, and cardiovascular dialysis patients have the highest rate of inci-dences. IDH may lead to nausea, vomiting, or anxiety. In some severe cases, there will be accident like shock and death. Moreover, patients’ physical monitoring is easily interrupted, and medical quality is often compromised, as nurses in a hemodialysis center joggle many tasks at the time. Therefore, the purpose of this paper is to integrate the technologies of wireless sensor networks (WSN) technologies, global systems for mobile communications (GSMs), and a MySQL database to construct an automatic monitoring system to collect physical data during hemodialysis. And an alert can be immediately sent to family members and nurses if IDH occurs. We then apply this wireless monitoring technology to the patients with IDH. This system is applied to an indoor space. The system framework includes two parts: the wireless networks and the controlling platform. During hemodialysis simulations, the average success ratio of physical data transmitting is 94.84%, the average reliability of systolic pressure is 1, of diastolic pressure is 0.89, and of pulse is 0.94, where the testing cases are collected from 6 persons, and the time interval is 2 hours. The loss ratio of physical data results from space constraints and data collision. |
URI: | http://tdr.lib.ntu.edu.tw/jspui/handle/123456789/43616 |
全文授權: | 有償授權 |
顯示於系所單位: | 生物機電工程學系 |
文件中的檔案:
檔案 | 大小 | 格式 | |
---|---|---|---|
ntu-98-1.pdf 目前未授權公開取用 | 3.48 MB | Adobe PDF |
系統中的文件,除了特別指名其著作權條款之外,均受到著作權保護,並且保留所有的權利。